DK800

Variable area flowmeter with glass cone and 125 mm / 4.92″ installation length

  • For flows ≥2.5 l/h or 0.65 GPH
  • MIN/MAX flow monitoring (SIL 2 conform acc. to IEC 61508-2)
  • Integrated needle valve
  • Measuring uncertainty: ±2.5% (acc. to VDI/VDE 3513, sheet 2)
  • Optional limit switches, flow regulator, panel mount
DK800

Overview

The DK46/47/48/800 is a variable area (VA) flowmeter for monitoring and adjusting extremely small flow rates of liquids (from ≥0.04 l/h or ≥0.1 GPH) and gases (from ≥5 Nl/h or 0.22 SCFH). It is thus particularly useful for sample flow monitoring in analyser systems and gas chromatographs or measuring liquids and gases in laboratories and pilot installations. Blinds and supports are available for panel mounting to purge panels or similar applications. By using a lab support, the flowmeters can be converted into a desktop unit.

The purgemeter operates without auxiliary power supply and is available with different glass tube and installation lengths from 90 to 325 mm / 3.55 to 12.8″. The medium is visible and the flow value can be read directly off the scale at the height of the float. The glass flowmeter can be provided with max. two limit switches (NAMUR) or a 3-wire digital output. For precise setting of the flow value, the VA meter features a needle valve. The DK 46/47/48/800 is also available with a differential pressure/flow regulator to provide constant flow rates in case of varying inlet or outlet pressures. The device has approvals for use in hazardous areas.

Product features and options depend on device configuration: Contact us to make sure that your preferred functionality can be combined in one device

  • Cost-effective small purgemeter
  • Low-cost measuring principle without auxiliary power
  • Maximum measuring error: ±1 / ±2.5 / ±4% (acc. to VDI/VDE 3513-2) or ±1 / ±2 / ±3% of full scale
  • Fine metering of very small volumes of liquids and gases
  • Integrated high quality needle valve in foot or head piece
  • Stainless steel, brass or PVDF fittings
  • Suitable for use in Safety Instrumented Systems with SIL 2, in redundant design (1oo2) also in SIL 3 applications (type A device)
  • No wearing parts

Various industries

  • Process or carrier gases
  • Purge fluids for measuring systems
  • Air or water
  • Nitrogen, CO2 or other inert gases
  • Sample flow monitoring in gas analysers etc.
  • Gas sparging in liquids

Conventional and nuclear power industry

  • Air, gases (nitrogen, hydrogen, methane etc.)
  • Water

Specifications

General information

Product typeFlowmeter
Product family nameDK
Product positioningFor low flow gas or liquid applications and sample flow monitoring
Measurement principleVariable area float
Measured mediaLiquids, Gases

Operating conditions

Process pressuremax. 10 barg/
145 psig
Process temperature0…+100°C/
-40…+392°F

Materials

Measuring tubeBorosilicate glass
GasketsEthylene propylene diene monomer (EPDM), Polytetrafluoroethylene (PTFE), Perfluoroelastomer rubber (FFKM), Fluoroelastomer rubber (FPM)
Process connectionAustenitic stainless steel (1.4404/316L), Polyvinylidene fluoride (PVDF), Nickel-plated brass

Process connections

Flange connectionsEN (1092-1): DN15…25
ASME (B 16.5): ½…1″
Thread connectionsFemale:
G¼ ¼ NPT
Other connectionsSwagelok;Serto, Ermeto; Hose; Gyrolok

Approvals/Standards

Certificates/Approvals ExATEX, NEPSI, PESO, UKEX
Certificates/Approvals NuclearNuclear qualification on request (e.g. acc. to ASME III, RCC-M/E, etc.)
Certificates/Approvals SafetySIL 2

Communication

Discrete outputsLimit switches
Scroll to Top